Shear stress for different pump types compared to typical range of shear stresses in blood vessels
Pump . | Examples . | Maximum shear stress (dynes/cm2) . |
---|---|---|
Typical range of wall shear stress in blood vessels17 | Large arteries | 14-36 |
Arterioles | 20-72 | |
Veins | 0.7-9 | |
Stenotic vessels | 36-450 | |
Positive displacement | Heartmate I | 100-5008-10,18 |
Thoratec PVAD | ||
Novacor N100 | ||
Rotary axial | Heartmate II | 2 000-24 00015,16,19,20 |
Jarvik 2000 | ||
Micromed DeBakey | ||
Rotary centrifugal | DuraHeart | 500-3 20015,16,21,22 |
Heartware VAD (HVAD) | ||
Levacor | ||
HeartMate 3 |
Pump . | Examples . | Maximum shear stress (dynes/cm2) . |
---|---|---|
Typical range of wall shear stress in blood vessels17 | Large arteries | 14-36 |
Arterioles | 20-72 | |
Veins | 0.7-9 | |
Stenotic vessels | 36-450 | |
Positive displacement | Heartmate I | 100-5008-10,18 |
Thoratec PVAD | ||
Novacor N100 | ||
Rotary axial | Heartmate II | 2 000-24 00015,16,19,20 |
Jarvik 2000 | ||
Micromed DeBakey | ||
Rotary centrifugal | DuraHeart | 500-3 20015,16,21,22 |
Heartware VAD (HVAD) | ||
Levacor | ||
HeartMate 3 |