Yearly mean and coefficient of variation (CV) values for the quality control (QC) samples of the Adelaide reference laboratory
Control . | QC target values . | 2001 . | 2002 . | 2003 . | 2004 . | 2005 . | 2006 (to March 28) . | New QC target values . | 2006 (from March 29) . | 2007 . |
---|---|---|---|---|---|---|---|---|---|---|
Low b2a2 | ||||||||||
Mean | 0.07 | 0.07 | 0.07 | 0.09 | 0.08 | 0.06 | 0.07 | 0.06 | 0.06 | 0.06 |
CV, % | 32 | 55 | 34 | 35 | 40 | 34 | 30 | 29 | ||
High b2a2 | ||||||||||
Mean | 56 | 48 | 53 | 69 | 50 | 42 | 52 | 22 | 21 | 20 |
CV, % | 33 | 33 | 22 | 25 | 28 | 29 | 26 | 29 | ||
Low b3a2 | ||||||||||
Mean | 0.07 | 0.07 | 0.06 | 0.08 | 0.06 | 0.08 | 0.09 | 0.08 | 0.08 | 0.08 |
CV, % | 54 | 52 | 37 | 41 | 35 | 34 | 34 | 35 | ||
High b3a2 | ||||||||||
Mean | 85 | 82 | 77 | 93 | 69 | 94 | 93 | 27 | 26 | 24 |
CV, % | 26 | 38 | 20 | 18 | 23 | 31 | 26 | 22 | ||
Number of RQ-PCR runs | 34 | 110 | 147 | 157 | 212 | 77 | 257 | 192 |
Control . | QC target values . | 2001 . | 2002 . | 2003 . | 2004 . | 2005 . | 2006 (to March 28) . | New QC target values . | 2006 (from March 29) . | 2007 . |
---|---|---|---|---|---|---|---|---|---|---|
Low b2a2 | ||||||||||
Mean | 0.07 | 0.07 | 0.07 | 0.09 | 0.08 | 0.06 | 0.07 | 0.06 | 0.06 | 0.06 |
CV, % | 32 | 55 | 34 | 35 | 40 | 34 | 30 | 29 | ||
High b2a2 | ||||||||||
Mean | 56 | 48 | 53 | 69 | 50 | 42 | 52 | 22 | 21 | 20 |
CV, % | 33 | 33 | 22 | 25 | 28 | 29 | 26 | 29 | ||
Low b3a2 | ||||||||||
Mean | 0.07 | 0.07 | 0.06 | 0.08 | 0.06 | 0.08 | 0.09 | 0.08 | 0.08 | 0.08 |
CV, % | 54 | 52 | 37 | 41 | 35 | 34 | 34 | 35 | ||
High b3a2 | ||||||||||
Mean | 85 | 82 | 77 | 93 | 69 | 94 | 93 | 27 | 26 | 24 |
CV, % | 26 | 38 | 20 | 18 | 23 | 31 | 26 | 22 | ||
Number of RQ-PCR runs | 34 | 110 | 147 | 157 | 212 | 77 | 257 | 192 |
All values were recorded whether the run was accepted or rejected. A new batch of QC samples was in use from March 2006.