Figure 1. Difference method for the measurement of the liver iron concentration by biomagnetic liver susceptometry: From the start position (left), the patient is lowered in an inhomogeneous magnetic field of 20–30 milliTesla by about 5 to 10 cm (right) to a field of 1 milliTesla. During this vertical scan, the water coupling bag follows the patient and the change of the magnetic flux as a function of the distance is registered in the detector coils and transformed by the SQUID electronics.
Figure 1.

Difference method for the measurement of the liver iron concentration by biomagnetic liver susceptometry: From the start position (left), the patient is lowered in an inhomogeneous magnetic field of 20–30 milliTesla by about 5 to 10 cm (right) to a field of 1 milliTesla. During this vertical scan, the water coupling bag follows the patient and the change of the magnetic flux as a function of the distance is registered in the detector coils and transformed by the SQUID electronics.